Title of article
A method of generalized projections (MGP) ghost correction algorithm for interleaved EPI
Author/Authors
K.J.، Lee, نويسنده , , N.G.، Papadakis, نويسنده , , D.C.، Barber, نويسنده , , I.D.، Wilkinson, نويسنده , , P.D.، Griffiths, نويسنده , , M.N.J.، Paley, نويسنده ,
Issue Information
روزنامه با شماره پیاپی سال 2004
Pages
-838
From page
839
To page
0
Abstract
Investigations into the method of generalized projections (MGP) as a ghost correction method for interleaved EPI are described. The technique is image-based and does not require additional reference scans. The algorithm was found to be more effective if a priori knowledge was incorporated to reduce the degrees of freedom, by modeling the ghosting as arising from a small number of phase offsets. In simulations with phase variation between consecutive shots for n-interleaved echo planar imaging (EPI), ghost reduction was achieved for n=2 only. With no phase variation between shots, ghost reduction was obtained with n up to 16. Incorporating a relaxation parameter was found to improve convergence. Dependence of convergence on the region of support was also investigated. A fully automatic version of the method was developed, using results from the simulations. When tested on invivo 2-, 16-, and 32-interleaved spin-echo EPI data, the method achieved deghosting and image restoration close to that obtained by both reference scan and odd/even filter correction, although some residual artifacts remained.