• Title of article

    Study of the surface contamination of copper with the improved positron annihilation-induced Auger electron spectrometer at NEPOMUC

  • Author/Authors

    J. Mayer، نويسنده , , C. Hugenschmidt، نويسنده , , K. Schreckenbach، نويسنده ,

  • Issue Information
    روزنامه با شماره پیاپی سال 2008
  • Pages
    3
  • From page
    220
  • To page
    222
  • Abstract
    The high intensity positron source NEPOMUC at the FRM-II in Munich enables measurement times for positron annihilation-induced Auger electron spectroscopy (PAES) of only 2.4 h/spectrum, in contrast to usual lab beams with measurement times up to several days. The high electron background due to surrounding experiments in the experimental hall of the FRM-II has been eliminated and hence background free experiments have become possible. Due to this, the signal to noise ratio has been enhanced to 4.5:1, compared to 1:3 with EAES. In addition, a long-term measurement has been performed in order to observe the contamination of a polycrystalline copper foil at 150 °C.
  • Keywords
    Auger electrons , Surface contamination , NEPOMUC , Positron annihilation
  • Journal title
    Applied Surface Science
  • Serial Year
    2008
  • Journal title
    Applied Surface Science
  • Record number

    1010189