Title of article
Fibre optic-based detection of the entire sample plug as a straightforward approach to kinetic measurements in flow-injection systems
Author/Authors
J.A. Garc?a-Mesa، نويسنده , , M.D.Luque de Castro، نويسنده , , M. Valc?rcel، نويسنده ,
Issue Information
روزنامه با شماره پیاپی سال 1994
Pages
4
From page
313
To page
316
Abstract
An unsegmented continuous-flow approach based on the use of tubular flow cells with an axial optical pathlength is reported. The device enables the monitoring of the entire sample plug and uses a spectrophotometric detector furnished with optical fibres. This allows the simple adaptation of the length of the glass tube that functions as the flow cell to the chemical system to be monitored. Recordings resembling those provided by the stopped-flow technique can be obtained without halting the flow, thus facilitating the implementation of reaction rate methods.
Keywords
Fibre optic , Flow systems , Reaction-rate measurements
Journal title
Analytica Chimica Acta
Serial Year
1994
Journal title
Analytica Chimica Acta
Record number
1022464
Link To Document