Title of article
Comparison of the analytical capabilities of an amperometric and an optical sensor for the determination of nitrate in river and well water
Author/Authors
Margaret A. Stanley، نويسنده , , Joe Maxwell، نويسنده , , Mairead Forrestal، نويسنده , , Andrew P. Doherty، نويسنده , , Brian D. MacCraith، نويسنده , , Dermot Diamond، نويسنده , , Johannes G. Vos، نويسنده ,
Issue Information
روزنامه با شماره پیاپی سال 1994
Pages
10
From page
81
To page
90
Abstract
The analysis of nitrate in water has been studied using novel amperometric and optical sensors. A flow-injection analysis system with amperometric detection has been developed in which nitrate is determined as nitrite after reduction in a cadmium column. The working electrode is glassy carbon modified with a crosslinked redox polymer. The linear range is 0.1 to 190 mgl NO3-N (r⪢0.999) and the limit of detection (LOD) is 50 μgl NO3-N. A fibre optic sensor based on a dual wavelength absorption approach has also been developed. A signal at 210 nm where nitrate absorbs was referenced against a signal at 275 nm where nitrate does not absorb. Its linear range is from 0.4 to 30 mgl NO3-N and its LOD is 400 μgl NO3-N. These diverse methods have been applied to the analysis of the same river water samples and good correlations have been observed between the two measurement techniques and a standard ion chromatography method.
Keywords
Amperometry , Sensors , Nitrate , Waters , Flow injection