Title of article
Development of a micro flow injection manifold for the determination of orthophosphate
Author/Authors
R.N.C. Daykin، نويسنده , , S.J. Haswell، نويسنده ,
Issue Information
روزنامه با شماره پیاپی سال 1995
Pages
5
From page
155
To page
159
Abstract
Fabrication of micro flow injection analysis (μFIA) systems based on glass substrates using lithographic techniques and etching methodology has been found to produce a reliable manifold for orthophosphate analysis based on colorimetric detection. The channel dimensions used were 30 μm deep and 325 μm wide, and mobility of the reagents and analytes was achieved by exploiting electrokinetic mobility (at pH < 3) and electroosmotic flow (at pH > 3). Detection was facilitated through the use of fibre optics coupled to a LED photodiode system. The time for analysis for the μFIA methods was found to be 3 min per analysis and the procedure had a limit of detection of 0.7 ppb (PO3−4) with R.S.D. values in the range 1–5%. The total volume of reactants in the μFIA device was estimated to be 0.5 μl.