    Time-based device used for the determination of tin by hydride generation flow injection atomic absorption techniques

  • Author/Authors

    M. Burguera، نويسنده , , J.L. Burguera، نويسنده , , C. Rivas، نويسنده , , P. Carrero، نويسنده , , R. Brunetto، نويسنده , , M. Gallignani، نويسنده ,

  • Issue Information
    روزنامه با شماره پیاپی سال 1995
  • Pages
    10
  • From page
    339
  • To page
    348
  • Abstract
    A time-based device used in conjunction with flow injection hydride generation atomic absorption spectrometry for the determination of tin in biological materials is described. Its configuration was designed for minimal manipulation of the samples with the great advantage of using large volumes of diluted digests. In order to ensure the determination of the analyte directly in the sample digest, several acids were tested for their effect on stannane generation. A mixture of nitric acid and perchloric acid was found to be suitable in an acidity range of 0.5–2.0 mol l−1; addition of hydrogen peroxide has no effect on the results. Other experimental parameters such as sample and reagent concentrations and volumes, carrier gas flow rate as well as atomizer conditions were optimized by both the simplex and univariate optimization methods. The results were in good agreement. A linear calibration graph was obtained in the range 5–250 μg l−1 with a 3σ detection limit of 2.5 μg l−1. The relative standard deviations were 4.8 and 2.6% for 5 and 50 μg l−1 tin, respectively. The method was applicable with good accuracy to the determination of tin in some biological tissues.
  • Keywords
    Biological materials , Hydride generation , Flow injection , Atomic absorption spectrometry , TIN
  • Journal title
    Analytica Chimica Acta
  • Serial Year
    1995
  • Journal title
    Analytica Chimica Acta
  • Record number

    1023605