Title of article
Fourier transform electrospray ion mobility spectrometry
Author/Authors
Yong Hong Chen، نويسنده , , William F. Siems، نويسنده , , Herbert H. Hill Jr.، نويسنده ,
Issue Information
روزنامه با شماره پیاپی سال 1996
Pages
10
From page
75
To page
84
Abstract
The analytical merit of Fourier transform electrospray ionization ion mobility spectrometry (FT-ESI-IMS) was investigated. When compared with signal-averaged electrospray ionization ion mobility spectrometry (SA-ESI-IMS) an increase in both signal/noise ratio (SN) and ion mobility resolving power was observed for the FT mode of operation. Using angiotensin II, glucosamine, n-tetradecylamine and gramicidin s as test compounds, electrospray ion mobility spectra were demonstrated in both modes under identical operating conditions. Both SN and IMS resolving power were measured as a function of gate pulse width. Although SN enhancement using FT was not as great as predicted from ideal conditions, a model was developed to explain this discrepancy. Not only did the FT operating mode enhance SN and IMS resolving power, it also served to evaluate the stability of the electrospray process.
Keywords
Fourier transform spectroscopy , Electrospray ionization , Ion mobility spectrometry
Journal title
Analytica Chimica Acta
Serial Year
1996
Journal title
Analytica Chimica Acta
Record number
1024298
Link To Document