Title of article
Glucose oxidase/hexokinase electrode for the determination of ATP
Author/Authors
Dario Compagnone، نويسنده , , George G. Guilbault، نويسنده ,
Issue Information
روزنامه با شماره پیاپی سال 1997
Pages
5
From page
109
To page
113
Abstract
A hydrogen peroxide based enzyme electrode for the determination of ATP has been developed by the immobilization of glucose oxidase and hexokinase. Competition between the enzymes for the substrate glucose allowed the measurement of ATP. Different immobilization procedures and different types of hexokinase have been tested. Using a BSA-glutaraldehyde procedure and hexokinase from an overproducing strain of bakersʹ yeast, ATP was measured in the 0.05–0.5 mmol l−1 range with a detection limit of 0.01 mmol l−1. ATP concentrations comparable to those reported in the literature and a good recovery were obtained when the enzyme electrode was used with human erythrocyte hemolysate.
Keywords
Glucose oxidase , ATP , Hexokinase , Amperometry , Biosensors
Journal title
Analytica Chimica Acta
Serial Year
1997
Journal title
Analytica Chimica Acta
Record number
1024421
Link To Document