Title of article
Limitations of quenching as a method of fluorometric analysis of non-fluorescent analytes
Author/Authors
Yener Rakicioglu، نويسنده , , Melissa Mickey Young، نويسنده , , Stephen G. Schulman، نويسنده ,
Issue Information
روزنامه با شماره پیاپی سال 1998
Pages
5
From page
269
To page
273
Abstract
It is shown that diffusional quenching of ordinary fluorophores (with decay times no longer than several tens of nanoseconds), a process employed in many fluorescence optical sensors, is rather insensitive as a analytical method having limits of detection no better than about 10−4 M. Longer lived luminophores, such as lanthanides, actinides and phosphorophores, however, can be quenched with substantially lower detection limits for the quencher. Static quenching, on the other hand, is limited by the thermodynamic strength of the complex(es) formed between the fluorophore and quencher rather than by the decay time of the fluorophore and the diffusibility of the quencher and can be quite sensitive as an analytical method for the quencher.
Journal title
Analytica Chimica Acta
Serial Year
1998
Journal title
Analytica Chimica Acta
Record number
1025471