Title of article
Determination of ultratrace levels of dissolved metals in seawater by reaction cell inductively coupled plasma mass spectrometry after ammonia induced magnesium hydroxide coprecipitation Original Research Article
Author/Authors
Francisco Ardini، نويسنده , , Emanuele Magi، نويسنده , , Marco Grotti، نويسنده ,
Issue Information
روزنامه با شماره پیاپی سال 2011
Pages
5
From page
84
To page
88
Abstract
A method for the determination of ultratrace amounts of Cr, Fe, Mn, Pb and Zn in seawater has been developed. It combined the low-blank magnesium hydroxide coprecipitation procedure with quadrupole inductively coupled plasma mass spectrometry and used the dynamic reaction cell technique to resolve the polyatomic interferences arising from the residual matrix, the solvent and plasma gases. Detection limits (3σB, n = 10) for Cr, Fe, Mn, Pb and Zn were 0.02, 0.10, 0.01, 0.002 and 0.19 nM, respectively, using 50 mL of seawater sample. The accuracy of the analytical procedure was verified by the analysis of the seawater reference materials CASS-4, NASS-5, SAFe D2 and SAFe S. The analytical precision ranged from 3% to 16% (n = 6), with a sample throughput of about 6 samples h−1.
Keywords
Trace elements , Inductively coupled plasma mass spectrometry , Polyatomic interferences , Matrix effects , Dynamic reaction cell , Seawater
Journal title
Analytica Chimica Acta
Serial Year
2011
Journal title
Analytica Chimica Acta
Record number
1026730