Title of article
Use of column with modified silica for interfering retention in a FIA spectrophotometric method for direct determination of vitamin C in medicine Original Research Article
Author/Authors
Julio Cesar B Fernandes، نويسنده , , Graciliano de Oliveira Neto، نويسنده , , Lauro T Kubota، نويسنده ,
Issue Information
روزنامه با شماره پیاپی سال 1998
Pages
12
From page
11
To page
22
Abstract
A flow injection system using a column packed with silica modified with aminopropyl groups as reactor in the spectrophotometric detection, at 265 nm, was developed for l-ascorbic acid determination. The column packed (2.79 i.d. and 50 mm length) with modified silica (200 mg) was used to eliminate interference by ionic pair formation between amine groups and interfering compounds. An analytical frequency of 30 samples per hour, a detection limit of 2.0×10−6 mol l−1 and a linear response range between 1.0×10−5 and 1.0×10−4 mol l−1 were obtained in optimized conditions. The estimation of relative standard deviation was 1.5% (n=20) for a 5.0×10−5 mol l−1 ascorbic acid solution. The method was applied in medicine samples to determine ascorbic acid and the results were in agreement of the nominal values.
Keywords
l-ascorbic acid determination , Medicine analysis , FIA , Modified silica
Journal title
Analytica Chimica Acta
Serial Year
1998
Journal title
Analytica Chimica Acta
Record number
1026890
Link To Document