• Title of article

    Air segmented continuous flow analysis in microbore tubing Original Research Article

  • Author/Authors

    J.L. Edwards Jr.، نويسنده , , R Bauman، نويسنده , , D.M Spence، نويسنده ,

  • Issue Information
    روزنامه با شماره پیاپی سال 1999
  • Pages
    6
  • From page
    209
  • To page
    214
  • Abstract
    Results from initial investigations of air-segmented continuous flow analysis (ASCFA) in microbore tubing are reported. Although excessive dispersion is limited in systems employing microbore tubing (i.e., <100 μm), the amount of mixing is increased from the recirculating flow pattern that forms due to the air bubbles. Response signals measured as a function of flow rate, reactor length and segmentation frequency show trends similar to ASCFA in conventional sized tubing. A comparison of systems with and without air segmentation is made employing a slow reaction, an enzyme catalyzed determination of glucose. Results show that the μ-ASCFA system produces a response signal that is 1.9 times the peak absorbance of the signal obtained without segmentation. In addition, the width of the response signal obtained with air segmentation is ∼50.0% of the width of the signal obtained without segmentation.
  • Keywords
    Continuous flow analysis , Capillary flow injection , pressure , Air segmentation
  • Journal title
    Analytica Chimica Acta
  • Serial Year
    1999
  • Journal title
    Analytica Chimica Acta
  • Record number