Title of article
An automatic titrator based on a multicommutated unsegmented flow system: Its application to acid–base titrations Original Research Article
Author/Authors
Cristina M.N.V. Almeida، نويسنده , , Rui A.S. Lapa، نويسنده , , José L.F.C. Lima، نويسنده , , Elias A.G. Zagatto، نويسنده , , M.C.U. Ara?jo، نويسنده ,
Issue Information
روزنامه با شماره پیاپی سال 2000
Pages
11
From page
213
To page
223
Abstract
A continuous flow methodology to perform acid–base titrations is described. The titrations are carried out on a multicommutated flow system which simulates batch titration procedures. The titration strategy is based on sequential insertion of increasing titrant and decreasing titrand volumes in a reactor, thus accomplishing complete titration curves. The assessment of the titration end point is similar to that of conventional batch procedures. The theoretical model for the determination of titrand concentration without requiring any calibration process is presented and discussed. The present system was evaluated in vinegar acidity determinations and provided an accuracy better than 3% with a good repeatability (relative standard deviation (RSD) = 2.5%; n = 10) and reproducibility (RSD < 5%). The titration accuracy is time-dependent and has been tested in the 2–10 min range.
Keywords
Multicommutated flow system , Titration , Acidity , Vinegar , Potentiometry
Journal title
Analytica Chimica Acta
Serial Year
2000
Journal title
Analytica Chimica Acta
Record number
1028835
Link To Document