• Title of article

    Performance characteristics for flow injection immunoassay using monoclonal antibodies against s-triazine and 2,4-D herbicides Original Research Article

  • Author/Authors

    Milan Franek، نويسنده , , Anping Deng، نويسنده , , Vladim??r Kol??، نويسنده ,

  • Issue Information
    روزنامه با شماره پیاپی سال 2000
  • Pages
    9
  • From page
    19
  • To page
    27
  • Abstract
    A sequential injection instrument (ALITEA, USA) with a photometric and fluorometric detection unit S2000 (Ocean Optics) was employed for the development of flow injection immunoanalysis (FIIA). The monoclonal antibodies against atrazine, simazine and 2,4-D were immobilized on aminopropyl glass particles by means of avidin/biotin system and packed in plexiglass column of 18 μl volume. Assay characteristics for individual antibody-reactors and regeneration effectivities for acid and alkaline solutions are described. An attempt to prepare a functional mixed antibody-reactor has not achieved success since regeneration conditions found for individual reactors were not compatible with one performance protocol.
  • Keywords
    Monoclonal antibodies , Immunoreactor , calibration , Flow injection immunoassay , Regeneration
  • Journal title
    Analytica Chimica Acta
  • Serial Year
    2000
  • Journal title
    Analytica Chimica Acta
  • Record number

    1029059