Title of article
Automatic determination of ascorbic acid by flame atomic absorption spectrometry Original Research Article
Author/Authors
M.C. Yebra-Biurrun، نويسنده , , R.M Cesp?n، نويسنده , , A Moreno-Cid، نويسنده ,
Issue Information
روزنامه با شماره پیاپی سال 2001
Pages
8
From page
157
To page
164
Abstract
A continuous flow procedure is proposed for the indirect determination of ascorbic acid based on its reducing properties. Chromium(VI) in acid medium was injected into a ascorbic acid stream, and the Cr(VI) was reduced to Cr(III). This Cr(III) formed was retained on-line, proportionally to the ascorbic acid concentration in the sample, on a poly(aminophosphonic acid) chelating resin, which is only selective for this oxidation state, while non-reduced Cr(VI) was determined by flame atomic absorption spectrometry. The proposed method allows the determination of ascorbic acid in the range 0.4–20 μg ml−1 with a relative standard deviation of 2.0% at a rate of ca. 90 samples per hour. This method has been applied to the determination of ascorbic acid in pharmaceutical preparations and fruit juices.
Keywords
Ascorbic acid , Flame atomic absorption spectrometry , Poly(aminophosphonic acid) chelating resin , Chromium , Flow injection
Journal title
Analytica Chimica Acta
Serial Year
2001
Journal title
Analytica Chimica Acta
Record number
1030760
Link To Document