Title of article
Determination of fluoxetine in plasma by gas chromatography–mass spectrometry using stir bar sorptive extraction Original Research Article
Author/Authors
Christian Fernandes، نويسنده , , Els Van Hoeck، نويسنده , , Pat Sandra، نويسنده , , Fernando M. Lanças، نويسنده ,
Issue Information
روزنامه با شماره پیاپی سال 2008
Pages
7
From page
201
To page
207
Abstract
This article presents a method employing stir bar sorptive extraction (SBSE) with in situ derivatization, in combination with either thermal or liquid desorption on-line coupled to gas chromatography–mass spectrometry for the analysis of fluoxetine in plasma samples. Ethyl chloroformate was employed as derivatizing agent producing symmetrical peaks. Parameters such as solvent polarity, time for analyte desorption, and extraction time, were evaluated. During the validation process, the developed method presented specificity, linearity (R2 > 0.99), precision (R.S.D. < 15%), and limits of quantification (LOQ) of 30 and 1.37 pg mL−1, when liquid and thermal desorption were employed, respectively. This simple and highly sensitive method showed to be adequate for the measurement of fluoxetine in typical and trace concentration levels.
Keywords
Thermal desorption , In situ derivatization , Plasma , Stir bar sorptive extraction , Fluoxetine
Journal title
Analytica Chimica Acta
Serial Year
2008
Journal title
Analytica Chimica Acta
Record number
1031590
Link To Document