Title of article
Screen-printed anodic stripping voltammetric sensor containing HgO for heavy metal analysis Original Research Article
Author/Authors
Jung-Yuen Choi، نويسنده , , Kyoungwon Seo، نويسنده , , Sung-Rok Cho، نويسنده , , Jae Ryoung Oh، نويسنده , , Sung-Hyun Kahng، نويسنده , , Jongman Park، نويسنده ,
Issue Information
روزنامه با شماره پیاپی سال 2001
Pages
7
From page
241
To page
247
Abstract
A screen-printed disposable sensor for the anodic stripping voltammetric (ASV) analysis of heavy metals was prepared by integrating a HgO-modified working electrode, Ag/AgCl reference electrode, and carbon counter electrode on a polycarbonate strip. The working electrode was screen-printed with phenol resin-based carbon ink containing fine particles of mercuric oxide as a built-in mercury precursor. The mercuric oxide particles exposed on the surface were reduced to fine mercury droplets by in situ or pre-cathodic conditioning in order to allow them to behave as heavy metal collectors in the anodic stripping analysis. The reference and counter electrodes were also screen-printed using commercial silver paste ink modified with finely ground AgCl powder and commercial carbon ink, respectively. The analytical characteristics of the sensor were evaluated using Pb and Cd as probe metals. Detailed preparation procedures and analytical characteristics of the sensor are described.
Keywords
Anodic stripping voltammetric analysis , heavy metals , Screen-printed disposable sensor
Journal title
Analytica Chimica Acta
Serial Year
2001
Journal title
Analytica Chimica Acta
Record number
1032638
Link To Document