• Title of article

    Determination of triazine herbicides by capillary liquid chromatography with on-column focusing and temperature gradient Original Research Article

  • Author/Authors

    M.E Le?n-Gonz?lez، نويسنده , , L.V. Pérez-Arribas، نويسنده , , L.M Polo D??ez، نويسنده , , C Panis، نويسنده , , M.P San Andrés، نويسنده ,

  • Issue Information
    روزنامه با شماره پیاپی سال 2001
  • Pages
    6
  • From page
    29
  • To page
    34
  • Abstract
    The combination of capillary liquid chromatography, temperature gradient and on-column focusing with large injection volumes offers a powerful tool for the determination of very low concentrations of triazines and their degradation products when very polar degradation products are present. A temperature gradient from subambient temperatures up to 70°C allows separation of eight triazines in 24 min. The focusing step makes it possible to increase 50 times the sample volume that can be injected into a capillary column. Detection limits (3×noise) are improved by a factor between 50 and 125 to 9 ng l−1 for hydroxyatrazine and 45 ng l−1 for atrazine. The chromatographic method, combined with solid phase extraction, was applied to the determination of eight triazines in ground water.
  • Keywords
    Capillary liquid chromatography , Triazine herbicides , Column focusing , Temperature programming , Water
  • Journal title
    Analytica Chimica Acta
  • Serial Year
    2001
  • Journal title
    Analytica Chimica Acta
  • Record number

    1032680