Title of article
Accurate and precise lithium isotopic determinations of igneous rock samples using multi-collector inductively coupled plasma mass spectrometry Original Research Article
Author/Authors
Yoshiro Nishio، نويسنده , , Shun’ichi Nakai، نويسنده ,
Issue Information
روزنامه با شماره پیاپی سال 2002
Pages
11
From page
271
To page
281
Abstract
A second-generation multi-collector inductively coupled plasma mass spectrometry (MC-ICP-MS) was applied to lithium isotopic measurements. The high sensitivity of the ICP-MS enabled high precision (±0.82‰, 2σ) analyses with small amount of Li (∼45 ng Li). A single-step column separation protocol was established with which rapid purification of lithium from rock solutions can be carried out with reduced blank (<10 pg). The influence of potential sources of error for acquisition of lithium isotopic data introduced during the separation, such as matrix effects and isotopic fractionation due to incomplete recovery, were examined with an artificially mixed solution of a composition similar to that of basalt, which was doped with Li isotopic standard reagent. The examinations demonstrated that our protocol suffered from negligible isotopic fractionation.
Keywords
Lithium isotopes , Magnetic sector , Multi-collector , ICP-MS , Igneous rock , Standard
Journal title
Analytica Chimica Acta
Serial Year
2002
Journal title
Analytica Chimica Acta
Record number
1032885
Link To Document