• Title of article

    Chronocoulometric flow-injection analysis with solvent polymeric membrane ion sensors Original Research Article

  • Author/Authors

    Concepci?n S?nchez-Pedre?o، نويسنده , , Joaqu??n A. Ortu?o، نويسنده , , Jorge Hern?ndez، نويسنده ,

  • Issue Information
    روزنامه با شماره پیاپی سال 2002
  • Pages
    7
  • From page
    11
  • To page
    17
  • Abstract
    A method to carry out chronocoulometric measurements with solvent polymeric membrane ion sensors in flow-injection systems has been developed. For this, a double potential step was synchronised to the passage of the sample plug through the detector cell. A four-electrode potentiostat with ohmic drop compensation and a new flow-through cell to incorporate the four-electrode and the membrane were developed. A plasticized poly(vinyl chloride) (PVC) membrane containing TBATBP was used and the procedure was applied to the determination of tetraethyl-ammonium. The effect of the electrochemical and flow-injection variables was studied. In the selected conditions, a linear relationship between the quantity of electricity and tetraethyl-ammonium concentration was obtained in the range 5×10−7–5×10−5 M. The detection limit was 7×10−8 M. Good repeatability and between day reproducibility were obtained. The potential application to other quaternary ammonium ions including acetylcholine was also studied.
  • Keywords
    Chronocoulometry , Flow-injection analysis , Solvent polymeric membrane , Ion sensor , Tetraethyl-ammonium determination
  • Journal title
    Analytica Chimica Acta
  • Serial Year
    2002
  • Journal title
    Analytica Chimica Acta
  • Record number

    1032970