Title of article
New triiodomercurate-modified carbon paste electrode for the potentiometric determination of mercury Original Research Article
Author/Authors
M.N. Abbas، نويسنده , , G.A.E Mostafa، نويسنده ,
Issue Information
روزنامه با شماره پیاپی سال 2003
Pages
7
From page
329
To page
335
Abstract
A new tetrazolium–triiodomercurate-modified carbon paste electrode has been described for the sensitive and selective determination of mercury. The electrode shows a stable, near-Nernstian response for 1×10−3 to 6×10−6 M [HgI3]− at 25 °C over the pH range of 4.0–9.0, with an anionic slope of 55.5±0.4 mV. The lower detection limit is 4×10−6 M with a fast response time of 30–50 s. Selectivity coefficients of a number of interfering anions and iodo complexes of some metal ions have been estimated. The interference from many of the investigated ions is negligible. The determination of 1–200 μg/ml of mercury in aqueous solutions shows an average recovery of 98.5% and a mean relative standard deviation of 1.6% at 50.0 μg/ml. The direct determination of mercury in spiked wastewater, metal amalgams and dental alloy gave results that compare favorably with those obtained by the cold vapor atomic absorption spectrometric method. Potentiometric titration of mercury and phenylmercury acetate with standard potassium iodide has been monitored using the developed triiodomercurate–carbon paste electrode (CPE) as an end point indicator electrode.
Keywords
Mercury , Tetrazolium–teriiodomercurate , Modified carbon paste electrode , Potentiometry
Journal title
Analytica Chimica Acta
Serial Year
2003
Journal title
Analytica Chimica Acta
Record number
1033381
Link To Document