• Title of article

    Flow injection electrochemical enzyme immunoassay based on the use of an immunoelectrode strip integrate immunosorbent layer and a screen-printed carbon electrode Original Research Article

  • Author/Authors

    Qiang Gao، نويسنده , , Ying Ma، نويسنده , , Zhiliang Cheng، نويسنده , , Weidong Wang، نويسنده , , Xiurong Yang، نويسنده ,

  • Issue Information
    روزنامه با شماره پیاپی سال 2003
  • Pages
    10
  • From page
    61
  • To page
    70
  • Abstract
    A flow injection amperometric immunoassay system based on the use of screen-printed carbon electrode for the detection of mouse IgG was developed. An immunoelectrode strip, on which an immunosorbent layer and screen-printed carbon electrode were integrated, and a proposed flow cell have been fabricated. The characterization of the flow immunoassay system and parameters affecting the performance of the immunoassay system were studied and optimized. Amperometric detection at 0.0 V (versus Ag/AgCl) resulted in a linear detection range of 30–700 ng ml−1, with a detection limit of 3 ng ml−1. The signal variation among electrode strips prepared from variant batch did not exceed 8.5% (n=7) by measuring 0.5 μg ml−1 antigen standard solution.
  • Keywords
    Screen-printed carbon electrode , Flow injection , Enzyme immunoassay