• Title of article

    A new electrochemiluminescent detection system equipped with an electrically controlled heating cylindrical microelectrode Original Research Article

  • Author/Authors

    Zhenyu Lin، نويسنده , , Jianjun Sun، نويسنده , , Jinhua Chen، نويسنده , , Liang Guo، نويسنده , , GuoNan Chen، نويسنده ,

  • Issue Information
    روزنامه با شماره پیاپی سال 2006
  • Pages
    5
  • From page
    226
  • To page
    230
  • Abstract
    A new electrochemiluminescent (ECL) detection system equipped with an electrically controlled heating cylindrical microelectrode (HME) was developed in this paper. The cylindrical microelectrode made of platinum wire (25 μm in diameter, 6 mm in long) was used as the working electrode of the ECL detection system, the temperature of the electrode could be controlled electrically. The Ru(bpy)32+-ECL and Ru(bpy)32+–C2O42−-ECL systems were used to evaluate this ECL detection system. The detection limit for oxalate was found to be 3.0 × 10−4 mol/L when Te (temperature of the HME) was 22 °C, and found to be 3.0 × 10−6 mol/L at 80 °C, which indicates that the detection limit can be improved greatly at higher Te, based on which, it is possible to establish a more sensitive method for measurement of ECL by using a heated microelectrode.
  • Keywords
    Oxalate , Ru(bpy)32+ , Electrochemiluminescence , Heated cylindrical microelectrode
  • Journal title
    Analytica Chimica Acta
  • Serial Year
    2006
  • Journal title
    Analytica Chimica Acta
  • Record number

    1035770