• Title of article

    Finite element analysis of the behaviour of microvoids in the cement mantle of cemented hip stem: Static and dynamic analysis

  • Author/Authors

    M.M. Bouziane، نويسنده , , B. Bachir Bouiadjra، نويسنده , , S. Benbarek، نويسنده , , M.S.H. Tabeti، نويسنده , , T. Achour، نويسنده ,

  • Issue Information
    ماهنامه با شماره پیاپی سال 2010
  • Pages
    6
  • From page
    545
  • To page
    550
  • Abstract
    In this study, the three-dimensional finite element (FE) method is used to analyse the stress distribution around microcavities in the cement mantle of total hip arthoplasty (THA). Static and dynamic loading were analysed. The effect of the position of the microdefect on the stress distribution is also highlighted. The obtained results show that microcavity located in the proximal zone of the prosthesis is subject to higher stress field. The static loading generates higher stresses than dynamic one if the microcavity is located in the proximal and distal zones of the prosthesis. The inverse case is observed when the microcavity is located in the medial zone.
  • Journal title
    Materials and Design
  • Serial Year
    2010
  • Journal title
    Materials and Design
  • Record number

    1068630