Title of article
Failure analysis of cemented metal-on-metal total hip replacements from a single centre cohort
Author/Authors
Michael Bryant، نويسنده , , Michael Ward، نويسنده , , Richard Farrar، نويسنده , , Robert Freeman، نويسنده , , Ken Brummitt، نويسنده , , John Nolan، نويسنده , , Anne Neville، نويسنده ,
Issue Information
ماهنامه با شماره پیاپی سال 2013
Pages
8
From page
226
To page
233
Abstract
This paper presents the failure analysis of cemented metal on metal (MoM) total hip replacements (THR) from a single centre cohort. In this paper all retrieved femoral stems were implanted and revised at the Norfolk and Norwich University hospital by three specialist hip surgeons. All patients were subjected to the same combination of metallic implants, operative and follow up procedure. Between 1997 and 2008, 652 Ultima TPS™ femoral stems with the Ultima MoM articulation (DePuy International, Leeds) were implanted. At 2–9 years follow up, 90 revisions had been performed. Upon removal extensive fretting-corrosion of the cemented portions of the femoral stems was seen, resulting in high metal ion levels. In order to fully understand the factors resulting in failure of these cemented MoM THR optical light microscopy, Scanning Electron Microscopy with integrated Energy Dispersive X-ray spectroscopy (SEM/EDX) and X-ray Photo-electron spectroscopy (XPS) were utilised in order to assess the surface morphology and chemistry of retrieved MoM THR and surface chemistry of the films found on the surface. Gross slip, plastic deformation and directionality of the surface were extensively seen on the proximal surfaces of the retrievals. A more corrosive phenomenon was in the distal regions of the stem, demonstrating a seemingly inter-granular attack. SEM/EDX analysis demonstrated that black films present commonly seen on the highly worn regions of the cemented portions of the femoral stem were a complex mixture of chromium oxide, protein and amorphous carbon. XPS demonstrated films were Cr, O rich, mixed with organic material.
Keywords
Bone cement , Femoral stems , Fretting-corrosion , Retrieval study , Case Study
Journal title
Wear
Serial Year
2013
Journal title
Wear
Record number
1092720
Link To Document