• Title of article

    Source and magnitude of error in an inexpensive image-based water level measurement system

  • Author/Authors

    Troy E. Gilmore، نويسنده , , François Birgand، نويسنده , , Kenneth W. Chapman، نويسنده ,

  • Issue Information
    روزنامه با شماره پیاپی سال 2013
  • Pages
    9
  • From page
    178
  • To page
    186
  • Abstract
    Recent technological advances have opened the possibility to use webcams and images as part of the environmental monitoring arsenal. The potential sources and magnitude of uncertainties inherent to an image-based water level measurement system are evaluated in an experimental design in the laboratory. Sources of error investigated include image resolution, lighting effects, perspective, lens distortion and water meniscus. Image resolution and meniscus were found to weigh the most in the overall uncertainty of this system. Image distortion, although largely taken into account by the software developed, may also significantly add to uncertainty. Results suggest that “flat” images with little distortion are preferable. After correction for the water meniscus, images captured with a camera (12 mm or 16 mm focal lengths) positioned 4–7 m from the water level edge have the potential to yield water level measurements within ±3 mm when using this technique.
  • Keywords
    Water level , Water stage , Machine vision , Edge detection , Instrument comparison , Uncertainty
  • Journal title
    Journal of Hydrology
  • Serial Year
    2013
  • Journal title
    Journal of Hydrology
  • Record number