• Title of article

    Validation of an ion chromatographic method for the quantification of anions in water Original Research Article

  • Author/Authors

    Cristina Lopez-Moreno، نويسنده , , Isabel Viera-Alcaide، نويسنده , , Ana M. Urbano، نويسنده ,

  • Issue Information
    روزنامه با شماره پیاپی سال 2010
  • Pages
    6
  • From page
    111
  • To page
    116
  • Abstract
    In this work, the validation of the method for the determination of chloride, nitrate and sulphate in potable drinking water and non-treated water was performed. The instrumental setup used was an automated ion chromatograph equipped with chemical suppression system for anion analysis. The significance of this paper falls on the statistical study of the analytical method, where all the uncertainty contributions involved in the process have been considered. Parameters such as repeatability, reproducibility, precision, accuracy, uncertainty, limits of detection and quantification, range of measurement and calibration figures of merit, were determined in this paper.
  • Keywords
    Chromatography , Potable water , Continental water , validation , Quality , Uncertainty , Accreditation , Reproducibility , Repeatability
  • Journal title
    Desalination
  • Serial Year
    2010
  • Journal title
    Desalination
  • Record number