Title of article
Direct sector field ICP-MS determination of metal impurities in detonation nanodiamond Original Research Article
Author/Authors
Dimitar P. Mitev، نويسنده , , Ashley T. Townsend، نويسنده , , Brett Paull، نويسنده , , Pavel N. Nesterenko، نويسنده ,
Issue Information
روزنامه با شماره پیاپی سال 2013
Pages
9
From page
326
To page
334
Abstract
A simple, rapid, sensitive and accurate method for the determination of metal impurities in detonation nanodiamond (DND), based upon the direct aspiration of aqueous suspensions of nanodiamond into a sector field inductively coupled plasma mass spectrometer (ICP-MS) is proposed. Quantitative release, atomisation and ionisation of the nanodiamond coated and encased elements, within the plasma, and their detection free from spectral interference, was confirmed by comparing results obtained from the direct ICP-MS analysis of aqueous suspensions with results from the ICP-MS analysis of the acidic digests of pre-ashed DND, obtained through DND combustion in air at 723 K for 48 h. The developed direct method was capable of the quantification of more than 30 elements at concentration levels of 10−8 wt.%, with acceptable levels of precision and accuracy, from an aqueous suspension of just 0.1 mg mL−1 DND.
Journal title
Carbon
Serial Year
2013
Journal title
Carbon
Record number
1125060
Link To Document