• Title of article

    Comparative performance of a flexible fixation implant to a rigid implant in static and repetitive incremental loading

  • Author/Authors

    W. McCartney، نويسنده , , B.J. Mac Donald، نويسنده , , A.G. Olabi and M.S.J. Hashmi، نويسنده ,

  • Issue Information
    روزنامه با شماره پیاپی سال 2005
  • Pages
    9
  • From page
    476
  • To page
    484
  • Abstract
    Four point bending tests under static and physiological repetitive loading were performed on two implant systems, one a rigid established plate system and the other a new flexible pin screw system. Both systems were made from stainless steel components. Under static loading the plate system had a higher ultimate yield and modulus of elasticity than the flexible pin screw system. Small repetitive incremental loading produced different results and showed that the flexible pin screw system had nearly always a higher ultimate yield point and a lower modulus of elasticity. This difference between the non-physiological static loads and the physiological repetitive loads indicates that the flexible pin screw system has the potential in certain situations to create more load sharing with the bone and hence a better fracture fixation system than the established plates and screws technique.
  • Keywords
    Fracture fixation , Orthopaedic , Stress analysis , design
  • Journal title
    Journal of Materials Processing Technology
  • Serial Year
    2005
  • Journal title
    Journal of Materials Processing Technology
  • Record number

    1179757