    Experimental verification of calculated radiation-induced heat conduction effects in the water absorbed dose calorimeter

  • Author/Authors

    Daniel A. Krauss، نويسنده ,

  • Issue Information
    دوهفته نامه با شماره پیاپی سال 2002
  • Pages
    9
  • From page
    99
  • To page
    107
  • Abstract
    Water absorbed dose calorimeters will be used as PTB standards for the dosimetry in radiation therapy. In this investigation, heat conduction effects in the calorimeter due to the irradiation of the non-water materials of the plane-parallel detector are investigated experimentally and by model calculations. It is shown that the effects during 60Co-γ-irradiation can be adequately described by finite-element simulations and that corresponding corrections in the order of 0.2% result for the typical operating mode of the calorimeter. Correction factors kC are determined for a variety of different irradiation conditions.
  • Keywords
    Heat conduction , Correction factor (kC) , Finite-element method , Water absorbed dose calorimeter , Plane-parallel detector cylinder
  • Journal title
    Thermochimica Acta
  • Serial Year
    2002