• Title of article

    Survey of the effect of fill volume on the values for the enthalpy and rate constant derived from isothermal microcalorimetry: applications of a newly developed test reaction

  • Author/Authors

    M.A.A. O’Neill، نويسنده , , A.E Beezer، نويسنده , , R.M. Deal، نويسنده , , A.C. Morris، نويسنده , , J.C. Mitchell، نويسنده , , J.A. Orchard، نويسنده , , J.A. Connor، نويسنده ,

  • Issue Information
    دوهفته نامه با شماره پیاپی سال 2003
  • Pages
    7
  • From page
    163
  • To page
    169
  • Abstract
    A recent paper has reported the results of an inter/intralaboratory study into a test and reference reaction for isothermal microcalorimeters, the imidazole catalysed hydrolysis of triacetin. The values derived were 2.8×10−6±9.7×10−8 dm3 mol−1 s−1 and −91.7±3 kJ mol−1 for the rate constant and enthalpy, respectively. This paper reports the findings of an investigation, using this test reaction, into the effect of ampoule fill volume on the recovery of the derived rate constant and enthalpy. Fill volume is likely to be of significance for all commercially available, heat conduction, isothermal microcalorimeters. Experiments, reported here, were performed in a thermal activity monitor (TAM, thermometric AB, Jarfalla, Sweden) operating in the batch mode, for 3, 4 and 20 ml glass ampoules and 4 ml stainless steel ampoules. It is shown that the rate constant can be accurately recovered at fill volumes greater than 50% but below 50% fill volume this accuracy diminishes.
  • Keywords
    Volume study , Triacetin , Isothermal microcalorimetry
  • Journal title
    Thermochimica Acta
  • Serial Year
    2003
  • Journal title
    Thermochimica Acta
  • Record number