• Title of article

    The construction and evaluation of a high pressure manifold and vessels for a Calvet type microcalorimeter

  • Author/Authors

    D.E.G. Jones، نويسنده , , A.-M. Turcotte، نويسنده , , R.C. Fouchard، نويسنده ,

  • Issue Information
    دوهفته نامه با شماره پیاپی سال 2003
  • Pages
    13
  • From page
    63
  • To page
    75
  • Abstract
    A Setaram C-80 calorimeter has been modified in order to measure the heat flow of energetic materials at pressures up to 69 MPa. A manifold and sample cells capable of operating at high pressure were designed, constructed and evaluated. This paper will describe, in detail, the high pressure manifold construction, safety assessment and calibration. As well, the results for initial trials with ammonium nitrate (AN), and pentaerythritol tetranitrate (PETN) at various pressures and heating rates will be discussed.
  • Keywords
    evaluation , High pressure manifold , Calvet microcalorimeter , Ammonium nitrate , PETN
  • Journal title
    Thermochimica Acta
  • Serial Year
    2003
  • Journal title
    Thermochimica Acta
  • Record number

    1196004