• Title of article

    A new general-purpose isothermal microcalorimeter for use at temperatures up to 200 °C

  • Author/Authors

    Ingemar Wads?، نويسنده , , Lars Wads?، نويسنده ,

  • Issue Information
    دوهفته نامه با شماره پیاپی سال 2003
  • Pages
    6
  • From page
    15
  • To page
    20
  • Abstract
    The design and some properties of a new general-purpose isothermal microcalorimeter are reported. The instrument is a twin thermopile heat conduction calorimeter, which is designed for use up to 200 °C. The calorimetric units and surrounding heat sink are suspended inside a hollow aluminium construction, which is thermostated. Above that unit a second thermostated block is positioned and the whole assembly is suspended inside a Dewar vessel. When the instrument is used at room temperature and below, the thermostated units are cooled by use of an insertion Peltier effect cooler. The instrument can be used with a wide range of different reaction vessels (diameter 14 mm). Baseline experiments have been conducted in the temperature range 15–200 °C. Typical values obtained during 10 h periods at 200 °C are ±3 and ±10 nW for the baseline drift and baseline fluctuations, respectively. The heat detection limit, determined by release of electrical energy, is about 2 μJ. Preliminary stability measurements have been conducted at 100 °C on samples of stabilised and non-stabilised polyamide film.
  • Keywords
    200 °C , Stability measurements , Isothermal microcalorimeter , Nanowatt
  • Journal title
    Thermochimica Acta
  • Serial Year
    2003
  • Journal title
    Thermochimica Acta
  • Record number

    1196115