Title of article
Simple methods for calibrating IR in TGA/IR analyses
Author/Authors
T.L. Slager، نويسنده , , F.M. Prozonic، نويسنده ,
Issue Information
دوهفته نامه با شماره پیاپی سال 2005
Pages
7
From page
93
To page
99
Abstract
This work reports the calibration of the IR signals from TGA effluent gas using decomposition of solids and evaporation of liquids. The IR signal separates the TGA weight losses for quantitation, and allows a successful deconvolution of the TGA data where several compounds are evolving from the TGA at the same time. It is unnecessary to heat the transfer lines and IR cell as long as the effluent has some volatility at ambient temperature. A DSC pinhole pan makes a very effective device for the calibration of low boiling components and produces calibration data equivalent to that achieved by decomposition of solid substrates. The value of the technique is demonstrated by quantitation of overlapping components in the TGA effluent.
Keywords
IR , TGA , calibration , Evolved gas analysis
Journal title
Thermochimica Acta
Serial Year
2005
Journal title
Thermochimica Acta
Record number
1196662
Link To Document