Title of article
Quenching of polymer inside aluminum DSC pans: Origin of an apparent artifact
Author/Authors
Huipeng Chen، نويسنده , , Peggy Cebe.، نويسنده ,
Issue Information
دوهفته نامه با شماره پیاپی سال 2008
Pages
3
From page
63
To page
65
Abstract
To obtain amorphous polymer samples for thermal analysis by differential scanning calorimetry (DSC), a typical procedure involves quenching the sample, which is encapsulated inside an aluminum DSC pan and lid, into liquid nitrogen or liquid nitrogen-cooled glass beads. We demonstrate that this procedure can cause an artifact endotherm in the subsequent DSC scan of the polymer. The artifact is not related to the polymer sample itself, but is possibly due to phase change of surface moisture introduced during the process of transferring into the DSC cell. A procedure to eliminate the artifact is discussed.
Keywords
Condensed moisture , Polymer glass transition , PBT , Nylon , Liquid nitrogen quenching , Aluminum DSC pan
Journal title
Thermochimica Acta
Serial Year
2008
Journal title
Thermochimica Acta
Record number
1198302
Link To Document