Title of article
Temperature calibration and electrical characterization of the differential scanning calorimeter chip UFS1 for the Mettler-Toledo Flash DSC 1
Author/Authors
E. Iervolino، نويسنده , , A.W. van Herwaarden، نويسنده , , F.G. van Herwaarden، نويسنده , , E. van de Kerkhof، نويسنده , , P.P.W. van Grinsven، نويسنده , , A.C.H.I. Leenaers، نويسنده , , V.B.F. Mathot، نويسنده , , P.J. French and P.M. Sarro، نويسنده ,
Issue Information
دوهفته نامه با شماره پیاپی سال 2011
Pages
7
From page
53
To page
59
Abstract
This paper reports on the temperature calibration and electrical characterization of the calorimeter chip UFS1 (internal design XI-400) developed for the new commercially available differential scanning calorimeter (DSC), the Flash DSC 1 of Mettler-Toledo. The chip consists of 2 identical membranes both with a p-type polysilicon microheater in the center of the membrane and a p/n-type polysilicon thermopile for measuring the sample temperature. The temperature calibration of the XI-400 is performed in the temperature range from 208 K to 723 K. An isothermal calibration is first performed to calibrate the heater resistance and the obtained curve is used to calibrate the integrated thermopile. The accuracy of the calibration is then determined by measuring the extrapolated onset temperature (Te) of primary standards. A detailed electrical characterization of the device is also reported. The calibration method implemented and the good temperature reproducibility of the device allow to use devices with uncalibrated heater resistance in the temperature range from 208 K to 723 K with a typical maximum error of ±5 K.
Keywords
Temperature calibration , Microheater , Calorimeter chip , Thermopile
Journal title
Thermochimica Acta
Serial Year
2011