Title of article
Frequency characteristic of an aspiration integrating small ion counter with a shielded collector
Author/Authors
Zygmunt Grabarczyk، نويسنده ,
Issue Information
روزنامه با شماره پیاپی سال 2001
Pages
6
From page
284
To page
289
Abstract
In the paper the frequency response of the Gerdien tube with a shielded collector was derived. The frequency response of the tube was defined as the ratio of the complex amplitude of the collector current to the complex amplitude of ion concentration at the inlet surface K( jω)=Ic( jω)/n( jω). It was found as a system response to impulse ion concentration excitation. The shape sin( fτ)/( fτ) of the frequency response was obtained ( f is the frequency of time varying ion concentration and τ the time of the flight of ions from the inlet to the collector). The results can be useful for designing ion counters with a modulated ion stream.
Keywords
Air-ions , Gerdientube , Transferfunction
Journal title
JOURNAL OF ELECTROSTATICS
Serial Year
2001
Journal title
JOURNAL OF ELECTROSTATICS
Record number
1264325
Link To Document