• Title of article

    Application of finite element analysis to the design of tissue leaflets for a percutaneous aortic valve

  • Author/Authors

    Smuts، نويسنده , , A.N. and Blaine، نويسنده , , D.C. and Scheffer، نويسنده , , C. and Weich، نويسنده , , H. and Doubell، نويسنده , , A.F. and Dellimore، نويسنده , , K.H.، نويسنده ,

  • Issue Information
    ماهنامه با شماره پیاپی سال 2011
  • Pages
    14
  • From page
    85
  • To page
    98
  • Abstract
    Percutaneous Aortic Valve (PAV) replacement is an attractive alternative to open heart surgery, especially for patients considered to be poor surgical candidates. Despite this, PAV replacement still has its limitations and associated risks. Bioprosthetic heart valves still have poor long-term durability due to calcification and mechanical failure. In addition, the implantation procedure often presents novel challenges, including damage to the expandable stents and bioprosthetic leaflets. In this study, a simplified version of Fung’s elastic constitutive model for skin, developed by Sun and Sacks, was implemented using finite element analysis (FEA) and applied to the modelling of bovine and kangaroo pericardium. The FEA implementation was validated by simulating biaxial tests and by comparing the results with experimental data. Concepts for different PAV geometries were developed by incorporating valve design and performance parameters, along with stent constraints. The influence of effects such as different leaflet material, material orientation and abnormal valve dilation on the valve function was investigated. The stress distribution across the valve leaflet was also examined to determine the appropriate fibre direction for the leaflet. The simulated attachment forces were compared with suture tearing tests performed on the pericardium to evaluate suture density. It is concluded that kangaroo pericardium is suitable for PAV applications, and superior to bovine pericardium, due to its lower thickness and greater extensibility.
  • Keywords
    Kangaroo pericardium , bovine pericardium , Percutaneous aortic valve , Finite element analysis , Biaxial tissue testing , Aortic valve leaflet , Constitutive model for skin
  • Journal title
    Journal of the Mechanical Behavior of Biomedical Materials
  • Serial Year
    2011
  • Journal title
    Journal of the Mechanical Behavior of Biomedical Materials
  • Record number

    1404638