Title of article
Simultaneous determination of oxygen transport characteristics of six membranes by hexagonal dissolved oxygen sensor system
Author/Authors
Kim، نويسنده , , Tai-Jin and Jurng، نويسنده , , Tae-Hoon and Chung، نويسنده , , Ui-Hong and Hong، نويسنده , , Suk-In، نويسنده ,
Issue Information
روزنامه با شماره پیاپی سال 2001
Pages
10
From page
11
To page
20
Abstract
In the present study, a hexagonal sensor system composed of a six port measurement device and six different cathodes with one common anode was developed for simultaneous determination of oxygen permeabilities or diffusivities of six different membranes. To minimize experimental variations caused by geometrical differences among six individual electrodes, the current output of each electrode was divided by individual correction factors determined by standardization with a typical membrane of fluorinated ethylene propylene (FEP). The hexagonal sensor system was then used to simultaneously determine the oxygen transport characteristics such as permeabilities or diffusivities of six different polypropylene membranes. The present study showed that a newly developed hexagonal dissolved oxygen sensor system can be used for simultaneous measurement of oxygen transport characteristics of six different membranes.
Keywords
MEMBRANE , Permeability , Simultaneous , Oxygen , Sensor , Diffusivity
Journal title
Sensors and Actuators B: Chemical
Serial Year
2001
Journal title
Sensors and Actuators B: Chemical
Record number
1412658
Link To Document