• Title of article

    Development and characterization of an ELISA assay in PDMS microfluidic channels

  • Author/Authors

    Eteshola، نويسنده , , E and Leckband، نويسنده , , D، نويسنده ,

  • Issue Information
    روزنامه با شماره پیاپی سال 2001
  • Pages
    5
  • From page
    129
  • To page
    133
  • Abstract
    This report describes the first preliminary evaluation of a heterogeneous sandwich enzyme-linked immunoassay in a PDMS microfluidic device. The PDMS devices were fabricated using replica molding against a patterned photoresist generated by photolithographic techniques. With this experimental setup, the microfluidic sensor chip was successfully used to quantify a model analyte (sheep IgM) with sensitivity down to 17 nM. The conventional blocking cocktail used in nearly all polystyrene microtiter plate-based ELISA assays failed to block the nonspecific adsorption of the analyte and secondary antibody. This report describes the successful use of surface modification chemistries and a modified blocking cocktail to reduce background due to nonspecific adsorption and to thereby increase the sensitivity of the assay. These protocols can be extended to the detection of a variety of analytes by immunoassay in PDMS microchannels.
  • Keywords
    Polydimethylsiloxane (PDMS) , flow-through , microfluidic device , Immunoassay , Sensor
  • Journal title
    Sensors and Actuators B: Chemical
  • Serial Year
    2001
  • Journal title
    Sensors and Actuators B: Chemical
  • Record number