Title of article
pH sensor based on deposited film of lead oxide on aluminum substrate electrode
Author/Authors
Eftekhari، نويسنده , , Ali، نويسنده ,
Issue Information
روزنامه با شماره پیاپی سال 2003
Pages
5
From page
234
To page
238
Abstract
Two modified electrodes were fabricated by deposition of thin films of α-PbO2 and β-PbO2 on aluminum substrate electrodes. Both electrodes exhibit potentiometric response to the hydrogen ion concentration and, consequently, can be used as pH sensors. The potentiometric responses of the modified electrodes were close to the theoretical slope of Nerustian response and comparable with glass electrodes. The electrochemical properties and stability of the constructed electrode were investigated. The possibility of the pH sensor for using in flow injection analysis (FIA) was also studied, and the optimum conditions were obtained.
Keywords
pH sensor , Nerustian slope , lead oxide , Solid film , Modified electrode , Al electrode
Journal title
Sensors and Actuators B: Chemical
Serial Year
2003
Journal title
Sensors and Actuators B: Chemical
Record number
1413105