• Title of article

    Quality control of chip manufacture and chip analysis using epoxy-chips as a model

  • Author/Authors

    Preininger، نويسنده , , C and Sauer، نويسنده , , U، نويسنده ,

  • Issue Information
    روزنامه با شماره پیاپی سال 2003
  • Pages
    6
  • From page
    98
  • To page
    103
  • Abstract
    Biochips are miniaturized, highly ordered analysis systems which offer the unique advantage of highly parallel analysis of thousands of analytes at the same time. Although this technique has been enthusiastically developed and has promised to improve and speed up numerous biological assays, the quality control of chip manufacture, chip analysis and data management has received less attention. llowing article compares three epoxy-containing chip surfaces (ARChip Epoxy, 3D-Link™, and EasySpot) with respect to their autofluorescence, immobilization capacity, background fluorescence and hybridization efficiency. Since data collected from biochip experiments are random snapshots with errors, inherently noisy and incomplete, we tried to evaluate technical factors causing variability and to set up quality control procedures for chip manufacture and chip analysis. Variabilities caused by arraying, glass substrate and polymer coating, fluorescent label and experimental conditions are discussed in details.
  • Keywords
    Biochip , quality control , Epoxy , Microarray
  • Journal title
    Sensors and Actuators B: Chemical
  • Serial Year
    2003
  • Journal title
    Sensors and Actuators B: Chemical
  • Record number

    1417225