Title of article
A study of the wear of explanted metal-on-metal resurfacing hip prostheses
Author/Authors
Joyce ، نويسنده , , Thomas J. and Langton، نويسنده , , David J. and Nargol، نويسنده , , Antoni V.F.، نويسنده ,
Issue Information
ماهنامه با شماره پیاپی سال 2011
Pages
6
From page
517
To page
522
Abstract
Due to their recent introduction there are few studies of retrieved resurfacing hip prostheses. Nine such components associated with groin pain in patients, and five associated with early fracture of the femur, were obtained and analysed using a roundness measuring machine. While the ‘fracture’ components showed no more than 3 μm out of roundness, components associated with groin pain showed between 15 and 92 μm out of roundness values. These latter results indicate wear and correlated with high metal ion levels in these patients, therefore the groin pain was likely associated with an adverse reaction to excessive metal wear debris.
Keywords
Failure , Metal wear , Wear measurement
Journal title
Tribology International
Serial Year
2011
Journal title
Tribology International
Record number
1426380
Link To Document