• Title of article

    A study of the wear of explanted metal-on-metal resurfacing hip prostheses

  • Author/Authors

    Joyce ، نويسنده , , Thomas J. and Langton، نويسنده , , David J. and Nargol، نويسنده , , Antoni V.F.، نويسنده ,

  • Issue Information
    ماهنامه با شماره پیاپی سال 2011
  • Pages
    6
  • From page
    517
  • To page
    522
  • Abstract
    Due to their recent introduction there are few studies of retrieved resurfacing hip prostheses. Nine such components associated with groin pain in patients, and five associated with early fracture of the femur, were obtained and analysed using a roundness measuring machine. While the ‘fracture’ components showed no more than 3 μm out of roundness, components associated with groin pain showed between 15 and 92 μm out of roundness values. These latter results indicate wear and correlated with high metal ion levels in these patients, therefore the groin pain was likely associated with an adverse reaction to excessive metal wear debris.
  • Keywords
    Failure , Metal wear , Wear measurement
  • Journal title
    Tribology International
  • Serial Year
    2011
  • Journal title
    Tribology International
  • Record number

    1426380