• Title of article

    Wear of cardiac lead outer insulation due to internal cable motion

  • Author/Authors

    Himes، نويسنده , , Adam M. Wilson، نويسنده , , Charles، نويسنده ,

  • Issue Information
    ماهنامه با شماره پیاپی سال 2013
  • Pages
    9
  • From page
    177
  • To page
    185
  • Abstract
    Insulation wear in implanted cardiac leads may result in serious failure modes, including inappropriate or missing therapy. A novel test apparatus intended to simulate a range of loading conditions between cardiac lead components has been developed. Wear testing performed in an aqueous environment at loads and velocities anticipated to occur in-vivo shows that a material with decreased toughness exhibited increased wear rate. Order of magnitude variability within test groups was observed, suggesting that reliability projections should incorporate the statistical nature of the wear rate. Ranking of implantable material wear rates are possible using this methodology.
  • Keywords
    Insulation , Leads , WEAR , Cardiac
  • Journal title
    Tribology International
  • Serial Year
    2013
  • Journal title
    Tribology International
  • Record number