• Title of article

    Effect of shelf versus accelerated aging of UHMWPE on delamination in knee wear simulation

  • Author/Authors

    Saikko، نويسنده , , Vesa، نويسنده ,

  • Issue Information
    ماهنامه با شماره پیاپی سال 2014
  • Pages
    7
  • From page
    10
  • To page
    16
  • Abstract
    Severe delamination wear, fracture and subsurface embrittlement caused by oxidation have been observed in gamma-irradiation-sterilized UHMWPE tibial components of knee prostheses. In the present study, gamma-air-sterilized, shelf aged UHMWPE disks delaminated and fractured within a few hours of testing in a ball-on-flat knee wear simulator. In an earlier study, gamma-air-sterilized disks subjected to accelerated aging showed only moderate adhesive wear in 8 week wear tests. The fact that established methods of accelerated aging did not cause clinically relevant damage should be noted because these methods are commonly used in the evaluation of new types of UHMWPE developed for clinical use.
  • Keywords
    Shelf aging , UHMWPE , Accelerated aging , Delamination wear
  • Journal title
    Tribology International
  • Serial Year
    2014
  • Journal title
    Tribology International
  • Record number