Title of article
Interaction between workpiece and CMM during geometrical quality control in non-standard thermal conditions
Author/Authors
Kruth، نويسنده , , Jean-Pierre and Vanherck، نويسنده , , Paul and Van den Bergh، نويسنده , , Christophe and Schacht، نويسنده , , Benny، نويسنده ,
Issue Information
فصلنامه با شماره پیاپی سال 2002
Pages
6
From page
93
To page
98
Abstract
Industrial quality demands have resulted in increasing attention towards the thermal behavior of coordinate measuring machines. The influence of the workpiece on the measurement accuracy is hereby often disregarded. This can lead to significant measurement errors. The described research examines the interaction between the measurement device and the measured object. Four distinct measurement error types that result from non-standard temperatures are listed. Temperature variations during the measurement lead to the most challenging situation. The key to measurement accuracy lies in linking the measurement time to the accompanying temperature variation. The possibilities of this methodology are indicated by an experiment on a reference object.
Keywords
Dimensional metrology , CMM , Differential expansion , Temperature variations , Measurement errors
Journal title
Precision Engineering
Serial Year
2002
Journal title
Precision Engineering
Record number
1428717
Link To Document