• Title of article

    Mercury(II)-selective polymeric membrane electrode based on the 3-[4-(dimethylamino)phenyl]-5-mercapto-1,5-diphenylpentanone

  • Author/Authors

    Yu، نويسنده , , Xiu and Zhou، نويسنده , , Zaide and Wang، نويسنده , , Yi and Liu، نويسنده , , Yan and Xie، نويسنده , , Qian and Xiao، نويسنده , , Dan، نويسنده ,

  • Issue Information
    روزنامه با شماره پیاپی سال 2007
  • Pages
    7
  • From page
    352
  • To page
    358
  • Abstract
    A new poly(vinylchloride) (PVC) membrane electrode based on 4-(4-N,N-dimethylphenyl)-2,6-diphenylpyrilium tetrafluoroborate as the ionophore was prepared, which shows a good selectivity for mercury(II) ions. The electrode indicates a good linear response for mercury(II) over a wide concentration range from 1.0 × 10−8 to 1.0 × 10−3 M, with a near-Nernstian slope of 34 mV per decade and the detection limit is 1.0 × 10−8 M. This membrane electrode operates well in the pH range of 2.5–7.0 and can be used for at least 30 days without observing any considerable deviation. Selectivity coefficients of some mono-, di- and trivalent metal ions were also determined. The sensor exhibits high selectivity towards mercury(II), compared with the alkali, alkaline earth and transition metal ions. The electrode was also applied as an indicator electrode for potentiometric titration of mercury(II) ion with I−.
  • Keywords
    Ion-selective electrode , mercury , ionophore , PVC Membrane , Potentiometry
  • Journal title
    Sensors and Actuators B: Chemical
  • Serial Year
    2007
  • Journal title
    Sensors and Actuators B: Chemical
  • Record number