• Title of article

    Simple and rapid amperometric monitoring of hydrogen peroxide in salivary samples of dentistry patients exploiting its electro-reduction on the modified/palladized aluminum electrode as an improved electrocatalyst

  • Author/Authors

    Pournaghi-Azar، نويسنده , , M.H. and Ahour، نويسنده , , F. and Pournaghi-Azar، نويسنده , , F.، نويسنده ,

  • Issue Information
    روزنامه با شماره پیاپی سال 2010
  • Pages
    6
  • From page
    334
  • To page
    339
  • Abstract
    This paper describes the use of an aluminum electrode plated by thin layer metallic palladium and modified by Prussian blue (PB/Pd–Al) for hydrodynamic amperometry of hydrogen peroxide (H2O2) exploiting its electrocatalytic reduction on the electrode. In optimum conditions the hydrodynamic amperometry was used for the determination of H2O2 at μM concentration level. It is found that the calibration graph is linear in the H2O2 concentration range 5 × 10−6 to 34 × 10−6 mol L−1 with correlation coefficient of 0.999. The detection limit of the method was about 4 × 10−6 mol L−1. The possibility of using the proposed method for selective determination of hydrogen peroxide in the presence of molecular oxygen is demonstrated. The method was successfully used for the monitoring of H2O2 in some salivary and environmental samples.
  • Keywords
    Modified aluminum electrode , Prussian blue , Hydrogen peroxide , Saliva analysis , Hydrodynamic amperometry
  • Journal title
    Sensors and Actuators B: Chemical
  • Serial Year
    2010
  • Journal title
    Sensors and Actuators B: Chemical
  • Record number