• Title of article

    Biofunctional structural design of SAW sensor chip surfaces in a microfluidic sensor system

  • Author/Authors

    Jung، نويسنده , , Adrian and Gronewold، نويسنده , , Thomas M.A. and Tewes، نويسنده , , Michael and Quandt، نويسنده , , Eckhard and Berlin، نويسنده , , Peter، نويسنده ,

  • Issue Information
    روزنامه با شماره پیاپی سال 2007
  • Pages
    7
  • From page
    46
  • To page
    52
  • Abstract
    Surface acoustic wave (SAW) sensor chips with a gold or SiO2-polymer surface were integrated into a microfluidic S-sens K5 sensor system to be modified by a novel procedure. Self-assembled composite monolayers were formed on the sensor chip surfaces, starting from special aminocellulose derivatives or from NH2-(organo-) polysiloxanes. Subsequently, the interface layers were biofunctionalized. In one assay, the enzyme glucose oxidase was immobilized on the aminocellulose surface. In a second assay, the formed composite monolayers became thrombin-sensitive by immobilization of an RNA anti-thrombin aptamer. The biofunctionalized surfaces did only display negligible amounts of non-specific protein binding with respect to elastase and bovine serum albumin (BSA). Thrombin was bound highly specific and with high sensitivity. It was possible to regenerate the surfaces under flow conditions.
  • Keywords
    Glucose oxidase , RNA anti-thrombin aptamer , Biofunctional modifications of SAW sensor chips , Aminocellulose monolayers , NH2-(organo-) polysiloxane composites
  • Journal title
    Sensors and Actuators B: Chemical
  • Serial Year
    2007
  • Journal title
    Sensors and Actuators B: Chemical
  • Record number

    1438903