Title of article
Determination of the total hardness in tap water using acoustic wave sensors
Author/Authors
Verيssimo، نويسنده , , Marta I.S. and Oliveira، نويسنده , , Joمo A.B.P. and Gomes، نويسنده , , M. Teresa S.R. Gomes?، نويسنده ,
Issue Information
روزنامه با شماره پیاپی سال 2007
Pages
5
From page
102
To page
106
Abstract
The determination of hardness in water is a useful parameter to control the quality of water for households and industrial uses. The present work proposes a new methodology as an alternative to the conventional EDTA titration. The determination of Ca2+ and Mg2+ is performed using two different coated piezoelectric quartz crystals. Poly(vinyl chloride) membranes, incorporating a plasticizer and a Mg or Ca ionophore were used to coat the piezoelectric quartz crystals.
s obtained by this new methodology shows that the tap water found in Portugal can be considered soft in Gouveia (9.4 ± 0.8 mg L−1 CaCO3) and in Vila Praia Âncora (15 ± 1 mg L−1 CaCO3), slightly hard in Leiria (59 ± 1 mg L−1 CaCO3), moderately hard in Aveiro (74 ± 1 mg L−1 CaCO3) and in Esposende (104 ± 1 mg L−1 CaCO3) and very hard in Tomar (225 ± 1 mg L−1 CaCO3). The results obtained by the proposed method are not significantly different (α = 0.05), both in terms of accuracy and precision, from the ones obtained by the conventional EDTA titration method. Besides providing individual concentration of Ca2+ and Mg2+, this new methodology also has the advantage of being less reagent, sample and time consuming.
Keywords
acoustic wave sensor , Magnesium , Calcium , Hardness
Journal title
Sensors and Actuators B: Chemical
Serial Year
2007
Journal title
Sensors and Actuators B: Chemical
Record number
1439149
Link To Document